Differential absorbance
\(\Delta\varepsilon\) in M⁻¹ cm⁻¹, \(c\) in mol L⁻¹, \(\ell\) in cm. Beer–Lambert applied separately to each circular component.
A chiral sample absorbs left- and right-circularly polarized light by slightly different amounts. That difference, usually 10−3–10−5 of the total absorbance, reports on absolute configuration, secondary structure, folding and binding.
Linear light is the sum of equal L and R components. Unequal absorption makes the sum elliptical (CD). Unequal refractive index rotates the ellipse (optical rotation, ORD). Real ΔA values are about a thousand times smaller than shown.
CD is reported in at least five interconvertible units. Most errors in the literature come from mixing them up or from a wrong concentration. The highlighted cards are the ones to memorize.
\(\Delta\varepsilon\) in M⁻¹ cm⁻¹, \(c\) in mol L⁻¹, \(\ell\) in cm. Beer–Lambert applied separately to each circular component.
Most spectropolarimeters display θ in millidegrees: \(\theta_{\text{mdeg}} = 32\,980\,\Delta A\).
Units deg cm² dmol⁻¹. \(N-1\) counts peptide bonds, the far-UV chromophores. Use this to compare proteins of different size.
\(c\) in mol L⁻¹, \(\ell\) in cm; units deg cm² dmol⁻¹. The factor 3298.2 = 100 · ln10 · 180 / (4π).
Electric-dipole-allowed π→π*: \(g\sim10^{-4}\text{–}10^{-3}\). Magnetic-dipole-allowed n→π* (ketones): \(g\sim10^{-2}\). Also used for CPL (\(g_{\text{lum}}\)).
A transition is CD-active only if its electric (\(\boldsymbol\mu\)) and magnetic (\(\boldsymbol m\)) transition dipoles are not perpendicular. In an achiral molecule \(R=0\) by symmetry.
Units esu² cm² (1 DBM = 0.9274×10⁻³⁸ esu² cm²). Compare R with TDDFT output after fitting the band.
How computed \(R\) values are broadened into a spectrum; \(\Delta\tilde\nu\) is the half-width at 1/e.
Absorption (CD) and dispersion (ORD) are two faces of one complex response. A single CD band produces an anomalous, sign-changing ORD curve centered on it.
Two states split by \(2V_{12}\) with equal and opposite rotational strengths give a bisignate couplet. Sign follows the helicity of the dipole pair (sign convention depends on how \(\mathbf R_{12}\) is defined). Try it below.
Luo & Baldwin (1997), \(T\) in °C, \(n\) residues. Suitable for helical peptides; use full deconvolution for globular proteins.
Aim for total absorbance (sample + buffer + cell) near 0.8–1.0 at the wavelength of interest. Above ~2 the photomultiplier HT voltage rises and data become unreliable.
Each window reports on a different structural level. Hover or tap a band for detail.
Far-UV (190–250 nm): peptide backbone n→π* (~220 nm) and π→π* (~190, 208 nm) transitions. Secondary structure.
To a first approximation a far-UV spectrum is a linear combination of basis spectra: \(\;[\theta](\lambda) = \sum_k f_k\,[\theta]_k(\lambda)\). Deconvolution algorithms invert this against reference sets of proteins with known structures. The basis curves here are idealized Gaussian models for teaching, not a reference set.
Signatures to recognize: α-helix gives a strong positive band near 192 nm and two minima at 208 and 222 nm. β-sheet gives a single minimum near 215–218 nm and a weaker positive band near 195 nm. Disordered chains give a deep minimum near 198 nm. A [θ]222/[θ]208 ratio above ~1.0 can indicate interacting helices such as coiled coils.
When two strong π→π* chromophores sit close together in space, their excited states mix. The result is two exciton states split by \(2V_{12}\) with opposite rotational strengths. The sign of the resulting bisignate Cotton effect gives the absolute sense of twist between the chromophores (Harada–Nakanishi). The model below uses two dipoles perpendicular to the connecting axis, twisted by dihedral angle α.
The couplet vanishes at α = 0° or 180° (no chirality) and at 90° (no coupling, so the two bands cancel). Coupling falls off as \(1/R^{3}\), so the amplitude drops quickly with distance. Here \(V_{12}\,[\text{cm}^{-1}] = 5.04\,\kappa\,\mu_1\mu_2/R^{3}\) with μ in debye and R in nm.
Modern instruments do not measure AL and AR separately. A photoelastic modulator switches the beam between L and R at ~50 kHz, and a lock-in amplifier extracts the tiny AC component riding on the DC transmitted intensity.
\(J_1\) is the first-order Bessel function; the PEM amplitude \(\delta_0\) is tracked with wavelength (quarter-wave retardation).
Ammonium d-10-camphorsulfonate checks both the magnitude scale and the far-UV response. Check it regularly and after lamp changes.
Most bad CD spectra come from absorbing buffers, wrong concentrations or saturated detectors, not from the instrument.
Convert the instrument's raw millidegrees into the units a paper needs. The example values describe a 0.2 mg mL⁻¹ lysozyme-sized protein in a 1 mm cell.
Following [θ]222 while heating gives a denaturation curve. For a reversible two-state transition the fraction folded follows from the Gibbs–Helmholtz equation. Check reversibility by cooling and re-scanning before fitting.
A large ΔCp curves the stability profile and can expose cold denaturation. In practice, fit sloping native and unfolded baselines as well. The van 't Hoff ΔH from CD equals the calorimetric ΔH only for a true two-state process.
| Tool | Approach | Use it for |
|---|---|---|
| DichroWeb | Web server running CONTIN-LL, SELCON3, CDSSTR, K2D with reference sets (SP175, SMP180 for membrane proteins) | Standard secondary-structure estimation with goodness-of-fit (NRMSD) |
| BeStSel | Eight structure classes including parallel/antiparallel and β-sheet twist | β-rich proteins, amyloid; fold recognition from CD |
| CDSSTR | Variable-selection SVD over reference proteins | Robust general estimates; needs data to ~190 nm or lower |
| K2D3 | Nearest-neighbour on theoretical spectra | Quick α/β estimate from 190–240 nm data |
| PDB2CD / DichroCalc | Predict CD from atomic coordinates | Checking a model or MD ensemble against experiment |
| TDDFT (Gaussian, ORCA, Turbomole) | Conformer search → Boltzmann-weighted computed R and ECD | Absolute configuration of small molecules and natural products |
Expect roughly ±5–10% accuracy for helix content, worse for β-sheet and turns. Report which algorithm and reference set you used, and the wavelength cut-off.
| Method | Measures | Range | Strength |
|---|---|---|---|
| ECD | ΔA for electronic transitions | 180–800 nm | Fast, low sample need, secondary structure |
| SRCD | ECD with synchrotron light | ≥ 160 nm | More transitions, better β/turn resolution, high-throughput |
| VCD | ΔA for vibrational transitions | 800–4000 cm⁻¹ | Absolute configuration without a chromophore; g ≈ 10⁻⁵ |
| ROA | Raman intensity difference for R vs L | Raman shift | Aqueous solutions, backbone and side-chain detail |
| ORD | Optical rotation vs wavelength | UV–vis | KK-related to CD; works away from absorption bands |
| MCD | CD induced by a magnetic field (A, B, C terms) | UV–NIR | Electronic structure of metal centres, even in achiral samples |
| CPL | Differential emission of L vs R light | emission | Chirality of excited states, glum |
| LD | A∥ − A⊥ for oriented samples | UV–vis | Orientation of transitions; a CD artefact in oriented films |